Electronic component search and free download site.
Transistors,MosFET ,Diode,Integrated circuits
Part Name  

E5052B View Datasheet(PDF) - HP => Agilent Technologies

Part NameE5052B HP
HP => Agilent Technologies HP
DescriptionSignal Source Analyzer 10 MHz to 7 GHz, 26.5 GHz, or 110 GHz
E5052B Datasheet PDF : 21 Pages
First Prev 11 12 13 14 15 16 17 18 19 20
E5053A Microwave Downconverter Specifications and General Information Summary
Table 10-3. E5053A front ports
Description
RF Input port
Input connector
Frequency range
Input level
Input damage level
LO outputs
Output connector
Output frequency
Frequency resolution
Output power
LO spurious
IF inputs
Input connector
Frequency range
Maximum input level
IF gain
Noise floor
Mixer bias current
Specification
APC-3.5 (female), 50 ohm nominal
3 GHz to 26.5 GHz
3 GHz to 10 GHz (fundamental mixing)
9 GHz to 26.5 GHz (third harmonics mixing)
< +10 dBm ( 3 GHz to 10 GHz band)
< +5 dBm (9 GHz to 26.5 GHz band)
> +23 dBm
SMA (female), 50 ohm nominal
3 GHz to 10 GHz
50 MHz
10 dBm to 16 dBm (3 GHz to 6 GHz)
10 dBm to 15 dBm (6 GHz to 10 GHz)
< –55 dBc (offset frequency > 300 Hz) typical
SMA (female), 50 ohm nominal
250 MHz to 1,250 MHz
0 dBm typical
0 dB to 35 dB in 5 dB step
< –163 dBm/Hz
–10 mA to +10 mA
Table 10-4. General information
Description
Supplemental information (nominal)
External reference signal input port
Input connector
Input frequency
Input level
BNC (female), 50 ohm nominal
10 MHz ± 10 Hz typical
-6 dBm to 6 dBm typical
Internal reference signal output port
Output connector
BNC (female), 50 ohm nominal
Output frequency
10 MHz ± 50 Hz typical
Output level
2.5 dBm ± 3 dB typical
USB port
type-B (female),
provides connection to E5052A/B
AC power Line (a third -wire ground is required)
AC frequency
47 Hz to 63 Hz
AC voltage
90 V to 132 V, or 198 V to 264 V
(automatically selected)
AC power
120 VA maximum
Table 10-5. Analyzer environmental and dimensions
Description
Supplemental information (nominal)
Operating environment
Temperature
Humidity
Altitude
Vibration
+10 degC to +40 degC
RH 20% to 80% at wet bulb temp.
< 29 degC (non-condensing)
0 to 2,000 m (0 to 6,561 feet)
0.21 G maximum, 5 Hz to 500 Hz
Non-operating storage environment
Temperature
–10 degC to +60 degC
Humidity
RH 20% to 90% at wet bulb temp.
< 40 degC (non-condensing)
Altitude
0 to 4,572 m (0 to 15,000 feet)
Vibration
0.5 G maximum, 5 Hz to 500 Hz
Instrument dimensions
See Figure 10-6, 10-7, 10-8.
Weight (NET)
11 kg