System Performance with the E5053A Microwave Downconverter – cont’d
About “mmW pplication”:
Phase noise measurements above 26.5 GHz can be done by using
external harmonic mixers (such as Agilent 11970 series) and a
power divider (splitter) with E5053A LO and IF terminals. The
E5052B’s mmW application software sets up appropriate
LO frequencies for the harmonic mixers.
Figure 10-4. System set-up for harmonic mixers (E5053A + E5052B)
Table 10-2-H. Frequency band example of phase noise
measurement with mmW harmonic mixers
Mixer model
Frequency band
N
11970A
26.5 to 40 GHz
8
11970Q
33 to 50 GHz
10
11970U
40 to 60 GHz
10
11970V
50 to 75 GHz
14
11970W
75 to 110 GHz
18