Electronic component search and free download site.
Transistors,MosFET ,Diode,Integrated circuits
Part Name  

E5052B View Datasheet(PDF) - HP => Agilent Technologies

Part NameE5052B HP
HP => Agilent Technologies HP
DescriptionSignal Source Analyzer 10 MHz to 7 GHz, 26.5 GHz, or 110 GHz
E5052B Datasheet PDF : 21 Pages
First Prev 11 12 13 14 15 16 17 18 19 20
Baseband Noise Measurement
Table 6-1. Baseband noise measurement performance
Description
Baseband input connector
Measurement frequency range
Measurement parameters
Measurement level range
Baseband input damage level
Noise floor level
Measurement uncertainty1
Measurement trigger
Specification
BNC, 50 ohm nominal, AC coupled
1 Hz to 100 MHz (E5052B)
10 Hz to 100 MHz (E5052B Option 011)
dBV/Hz, dBm/Hz, V/√Hz
< +5 dBm
> +23 dBm, > 35 V DC
See Table 6-2.
± 4 dB (< 1 kHz) SPD
± 2 dB (> 1 kHz) typical
continuous/single/hold source: internal/external/manual/bus
Table 6-2. Baseband noise floor [dBm/Hz]
correlation = 1, baseband input: 0 ohm terminated
BB noise floor
Baseband frequency [Hz]
1
10
100
1k
10 k
100 k
1M
E5052B start frequency = 1 Hz, measurement time = 13 s
specification
–151
–158
–163
–160
typical
–119
–132
–145
–155
–162
–167
–164
E5052B Option 011 start frequency = 10 Hz, measurement speed = 3.3 s
specification
–148
–155
–160
–157
typical
–129
–142
–152
–159
–164
–161
10 M
–160
–164
–160
–164
40 M 100 M
–156 –156
–160 –160
–156 –156
–160 –160
Internal Timebase
Table 7-1. Internal timebase (OCXO) performance
Description
Frequency uncertainty
Frequency temperature coefficient
Frequency aging rate
Specification
± 5 Hz at 10 MHz (± 0.5 ppm)
< 0.5 ppb/degC
< 0.5 ppb/day
24 hours after a cold start for < 30 days continuous operation
1. Baseband measurement uncertainty: specified at > –60 dBm level.
11
Direct download click here

 

Share Link : 
All Rights Reserved© datasheetq.com 2015 - 2019  ] [ Privacy Policy ] [ Request Datasheet  ] [ Contact Us ]