1 2 3 4 5 6 7 8 9 10 Next Last
Corrected System Performance
The specifications in this section apply to measurements made with the
Agilent E5071C network analyzer under the following conditions:
• No averaging applied to data
• Environmental temperature of 23 °C (±5 °C) with less than 1 °C deviation
from the calibration temperature
• Response and isolation calibration performed
System dynamic range
Table 1. Option 230/235/240/245/260/265/280/285/430/435/440/445/
460/465/480/485
Description
System dynamic range1, 2
Specification
SPD
9 kHz to 300 kHz
72 dB
300 kHz to 10 MHz
10 MHz to 6 GHz
IF bandwidth = 3 kHz
82 dB
98 dB
6 GHz to 8.5 GHz
92 dB
9 kHz to 300 kHz
97 dB
115 dB
300 kHz to 10 MHz
107 dB
115 dB
10 MHz to 6 GHz
6 GHz to 7 GHz
IF bandwidth = 10 Hz
123 dB
117 dB
130 dB
128 dB
7 GHz to 8 GHz
117 dB
126 dB
8 GHz to 8.5 GHz
117 dB
124 dB
160
150
140
130
120
110
100
90
80
1.E+05
1.E+09
2.E+09
E5071C specification
3.E+09 4.E+09 5.E+09 6.E+09 7.E+09 8.E+09
Test frequency [Hz]
Figure 1. System dynamic range (specification and actual measurement data example,
IF bandwidth 10 Hz)
1. The test port dynamic range is calculated as the difference between the test port rms noise floor
and the source maximum output power. The effective dynamic range must take measurement
uncertainty and interfering signals into account.